Metric large increment buttons update
The large increment buttons (currently +-5%) need to be updated to reflect intuitive use. 'Ease of use' buttons need to be added for a number of metrics as well. Changes need to be confirmed with Ramiro before merged back into master.
BPM:
- Change +- buttons to be +-1
- Needs buttons for range 6-28 BPM
I:E:
- Needs buttons allow favoring of inspiration. We currently have 1:1-4, but this should reflect 1-4:1-4. Eg. if 1:4 is possible, so should 4:1. CONFIRM RANGE AND BUTTONS WITH RAMIRO BEFORE IMPLEMENTING!
Tidal Volume:
- Change +-5% buttons to be +- 200mL
- Change +- buttons to be +- 50mL
- Fix range to reflect 150mL to 2,000ML
Minute Ventilation:
- Change +-5% buttons to +- 5L